Product Description
Relevant Standard Testing Requirements
1.1 The test object is a battery pack or system.
1.2 Conduct four experimental cycles with reference to the severity level (5) of GB/T2423.18.
1.3 The salt solution is prepared with sodium oxide (chemically pure, analytically pure) and distilled water or deionized water. Its concentration is (5 ± 0.1)% (mass fraction). Measure the pH value at (20 ± 2) °C, and the pH value should be between 6.5 and 7.2.
1.4 Put the test object into a salt fog chamber. Spray salt fog at a temperature between 5 °C and 35 °C for 2 hours. After the spraying is completed, transfer the test object to a damp heat chamber and store it for 20 to 22 hours at a temperature of (40 ± 2) °C and a relative humidity of (93 ± 3)%, which constitutes one cycle. Repeat this cycle three times, and then store the test object for 3 days under the standard atmospheric test conditions (temperature: 23 °C ± 2 °C, relative humidity: 45% to 55%), which constitutes one period. Repeat the test for four periods.
1.5 Requirements: There should be no leakage, shell rupture, fire or explosion in the battery pack or system.
![]()
Main Parameters
Name of the equipment |
Compound salt spray test chamber |
Model |
MHY-1600M |
MHY-012M |
Inner box size (W*H*D) mm |
1000*1000*1600 (Excluding the spire) |
3000*2000*2000 (Excluding the spire) |
Temperature range |
0ºC~60ºC |
Temperature fluctuation |
±0.5ºC (When it is empty) |
Temperature uniformity |
±2ºC (When there is no load) |
Salt water spray amount |
1 to 2mL/ eight cm² 1Hr (Collect for at least 16 hours,take its average) |
Spray pressure |
1kg/cm² It is adjusted in two sections, and the first time the compressed air adjust to 2kg/cm². Enter the saturated barrel for heating, the the second adjustment is 1kg/cm². |
Laboratory temperature range |
0ºC~60ºC |
Humidity accuracy |
Temperature resolution: 0.1ºC |
Humidity range |
20%-98%RH |
Temperature fluctuation |
±3%RH |
Liquid medicine PH Value |
Neutral 6.0~7.0 Acid 3.0~3.1 |
Spray solution PH Value |
Neutral 6.5~7.2 Acid 3.1~3.3 |
Time range |
1S~9999H can be set arbitrarily |
Control method |
Device host and remote PC control, device body and console. Separate, can be controlled remotely, and the operator should stay away from the test area. The longest long-distance distance can reach 50 meters, and the standard line length is 10m. It can effectively protect the safety of testers. |